Conformidade de segurança
Todas as entradas estão protegidas de acordo com a CE, IEN61010-1, CAT III 600 V.
DETAILED SPECIFICATIONS
Fluke 114, Digital Multimeters
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity of
0 % to 90 %.
General specifications |
|||||
Maximum voltage (between any terminal and earth ground) |
600 V rms |
||||
Safety |
IEC 61010-1, Pollution Degree 2 IEC 61010-2-033 CAT III 600 V, 10 A EMC IEC 61326-1: Portable |
||||
Fuse for A input (115, 117 only) |
11 A, 1000 V, IR 17 kA (Fluke PN 803293) |
||||
Display |
Digital: 6,000 counts, updates 4/sec; Bar Graph: 33 segments, updates 32/sec |
||||
Temperature |
|||||
Operating |
-10 °C to +50 °C |
||||
Storage |
-40 °C to +60 °C |
||||
Humidity |
0 % to 90 % to 35 °C; 75 % to 40 °C; 45 % to 50 °C |
||||
Temperature coefficient |
0.1 x (specified accuracy/°C) (< 18 °C or > 28 °C) |
||||
Operating altitude |
2,000 meters |
||||
Battery |
9 Volt Alkaline (IEC 6LR61) |
||||
Battery life |
400 hours typical, without backlight |
||||
Certifications |
CE, CSA, RCM |
||||
IP rating (dust and water protection) |
IP42 |
||||
Accuracy specifications (all models) |
|||||
Function |
Range |
Resolution |
Accuracy ± ([% of Reading]+ [Counts]) |
Model |
|
DC millivolts |
600.0 mV |
0.1 mV |
0.5 % + 2 |
114, 115, 116, 117 |
|
DC volts |
6.000 V 60.00 V 600.0 V |
0.001 V 0.01 V 0.1 V |
0.5 % + 2 |
114, 115, 116, 117 |
|
DC, 45 Hz to 500 Hz |
500 Hz to 1 kHz |
||||
Auto-V LoZ1 true-rms (114, 116, 117 only) |
600.0 V |
0.1 V |
2.0 % + 3 |
4.0 % + 3 |
114, 116, 117 |
45 Hz to 500 Hz |
500 Hz to 1 kHz |
||||
AC millivolts1 true-rms |
600.0 V |
0.1 V |
1.0 % + 3 |
2.0 % + 3 |
114, 115, 116, 117 |
AC volts1 true-rms |
6.000 V 60.00 V 600.0 V |
0.001 V 0.01 V 0.1 V |
1.0 % + 3 |
2.0 % + 3 |
114, 115, 116, 117 |
Continuity (115, 116 only) |
600 Ω |
1 Ω |
Beeper on < 20 Ω, off > 250 Ω; detects opens or shorts of 500 μs or longer |
114, 115, 116, 117 |
|
Ohms |
600.0 Ω 6.000 kΩ 60.00 kΩ 600.0 kΩ 6.000 MΩ 40.00 MΩ |
0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ |
0.9 % + 2 0.9 % + 1 0.9 % + 1 0.9 % + 1 0.9 % + 1 5.0 % + 2 |
114, 115, 116, 117 |
|
Diode test |
2.000 V |
0.001 V |
0.9 % + 2 |
115, 116, 117 |
|
Capacitance (115, 116, 117 only) |
1000 nF 10.00 μF 100.0 μF 9999 μF - |
1 nF 0.01 μF 0.1 μF 1 μF - |
1.9 % + 2 1.9 % + 2 1.9 % + 2 100 μF to 1000 μF: 1.9 % + 2 > 1000 μF: 5 % + 20 |
115, 116, 117 |
1 All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of range are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of
≤ 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of ≤ 3. AC volts is ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled.
Accuracy specifications (all models) |
||||
Function |
Range |
Resolution |
Accuracy ± ([% of Reading]+ [Counts]) |
Model |
LoZ capacitance (power-up option) (114, 116, 117 only) |
1 nF to 500 μF |
10 % + 2 typical |
115, 116, 117 |
|
Temperature - K-Type thermocouple (116 only) |
-40 °C to 400 °C -40 °F to 752 °F |
|
1 % + 102 1 % + 182 |
116 |
AC amps true-rms1 (45 Hz to 500 Hz) (115, 117 only) |
6.000 A 10.00 A 20 A overload for 30 seconds maximum |
0.001 A 0.01 A |
1.5 % + 3 |
115, 117 |
AC μAmps true-rms1 (45 Hz to 1 kHz) (116 only) |
600.0 μA |
0.1 μA |
1.5 % + 3 (2.5 % + 3 > 500 Hz) |
116 |
DC amps (115, 117 only) |
6.000 A 10.00 A 20 A overload for 30 seconds maximum |
0.001 A 0.01 A |
1.0 % + 3 |
115, 117 |
DC μAmps true-rms (116 only) |
600.0 μA |
0.1 μA |
1.0 % + 2 |
116 |
Hz (V or A input)2 (Hz A 115, 117 only) |
99.99 Hz 999.9 Hz 9.999 kHz 50.00 kHz |
0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz |
0.1 % + 2 |
115, 117 |
Hz (V input)3 (115, 116, 117 only) |
99.99 Hz 999.9 Hz 9.999 kHz 50.00 kHz |
0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz |
0.1 % + 2 |
116 |
All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of range are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of
≤ 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of ≤ 3. AC volts is ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled. 2 AC Volts Hz is ac-coupled and specified from 5 Hz to 50 kHz. AC Amps Hz is dc-coupled and specified from 45 Hz to 5 kHz. Amps input burden
voltage (typical): 6 A input 2 mV/A, 10 A input 37 mV/A.
3 Frequency is ac-coupled, 45 Hz to 50 kHz.
Frequency counter sensitivity (models 115, 116, 117) |
|||||
Input range |
Typical sensitivity (rms sine wave) |
||||
5 Hz to 45 Hz |
45 Hz to 5 kHz |
5 kHz to 10 kHz |
10 kHz to 50 kHz |
||
Volts AC |
V 60 V 600 V |
0.2 V 2 V 20 V |
0.2 V to 0.3 V 2 V to 3 V 20 V to 30 V |
0.3 V to 0.4 V 3 V to 4 V 30 V to 40 V |
0.4 V to 1.0 V 4 V to 10 V 40 V to 100 V |
AC Amps (115, 117 only) |
6 A 10 A |
N/A N/A |
0.4 A 0.5 A |
N/A N/A |
N/A N/A |
Input characteristics (all models) |
|||||
Function |
Input impedance (nominal) |
Common mode rejection ratio (1 kΩ unbalanced) |
Normal mode rejection |
||
Volts AC |
> 5 MΩ < 100 pF |
> 60 dB at dc, 50 or 60 Hz |
|||
Volts DC |
> 10 MΩ < 100 pF |
> 100 dB at dc, 50 or 60 Hz |
> 60 dB at 50 or 60 Hz |
||
Auto-V LoZ (114, 116, 117 only) |
~3 kΩ < 500 pF |
> 60 dB at dc, 50 or 60 Hz |
|||
Open circuit test voltage |
Full scale voltage |
Short circuit current |
|||
Ohms |
< 2.7 V dc |
To 6.0 MΩ |
40 MΩ |
< 350 μA |
|
< 0.7 V dc |
< 0.9 V dc |
||||
Diode test (115, 117 only) |
< 2.7 V dc |
2.000 V dc |
< 1.2 mA |